New TopShim run: version = topshim 2.1.0 date = 2021-12-23 time = 11:30:39.887 command = topshim tunea FINE SHIMMING Parameters: probe = Z168276_0016 solvent = H2O+DMSO shim nucleus = 1H nucleus optimised for = 1H o1p (set) = 4.70 ppm Results: 1D Shimming: Iteration 1: FWHM Before= 4.8 Hz, Target= 0.4 Hz, After= 1.0 Hz Iteration 2: FWHM Before= 1.0 Hz, Target= 0.4 Hz, After= 0.7 Hz Iteration 3: FWHM Before= 0.7 Hz, Target= 0.4 Hz, After= 0.3 Hz Iteration 4: FWHM Before= 0.3 Hz, Target= 0.4 Hz, After= 0.5 Hz Iteration 5: FWHM Before= 0.5 Hz, Target= 0.4 Hz, After= 0.4 Hz Shim succeeded Lineshape quality factor (1D spectrum, spinning) = 0.468 Performing tune after shimming tune lock level increase = 1.0 shim changes: Z +101 Z2 +22 Z3 +38 Z4 -175 Z5 -134 Z6 -72 Z7 -147 Z8 +6205 X +5 XZ +28 Y +62 YZ +33 duration = 1 min 44 sec completed successfully finished Thu Dec 23 11:32:22 2021 ========================================================================